User's Manual

66 US-ENGLISH
20.1. TABLE OF DELIVERED SHOCK ENERGY
AND VOLTAGE
The relationship between stored energies, maximum voltages and
delivered energies (at 37 °C, 50 ohm load) for the minimum, low, mean
and maximum programmed energy values is as follows:
Stored energy (J)
0.5
10
20
34
42
V1 (Volt)
75
341
483
631
702
V2 (Volt)
37
173
245
318
353
Delivered E: Phase 1 (J)
0.31
7.0
14.0
23.9
29.6
Delivered E: Phase 2 (J)
0.08
1.8
3.6
6.1
7.5
Delivered E: Total (J)
0.4
8.8
17.6
30.0
37.1
Tolerances are 12% for voltage (25% at 0.5 J) and 30% for energy.
20.2. BATTERY
Manufacturer
Greatbatch
Type
Quasar High Rate (QHR)
Model
GB 2593
Number of batteries
1
Total capacity
1964 mAh
Usable capacity
Between BOL and ERI: 1278 mAh.
Between BOL and EOL: 1675 mAh.
Voltage
BOL: 3.25 V. ERI: 2.66 V. EOL: 2.5 V.