User's Manual

56 US-ENGLISH
17.3. ELECTIVE REPLACEMENT INDICATOR (ERI)
Elective Replacement Indicators (ERI)
(1)
are:
magnet rate equal to 80 ± 1 min-1 or
battery voltage equal to 2.66 V ± 0.01 V
Caution: The defibrillator should be replaced as soon as the Elective
Replacement Indicator (ERI) point is reached.
Between the ERI and the EOL (End of Life)
(2)
, PARADYM RF CRT-D
9750 can still function for:
7.4 months (100% atrial and biventricular pacing in DDD mode,
500 ohms, with as-shipped settings), and deliver 7 shocks at 34 J
or
6.4 months (0% pacing, sensor OFF, one 42 J shock every
2 weeks).
Once the Elective Replacement Indicator (ERI) point has been
reached, the device operates normally, except that the charge time
increases. Under normal conditions (and without programmer use) the
charge times are as as follows:
(1) Elective Replacement Indicators (ERI) corresponds to Recommended
Replacement Time (RRT) as referred in the EN45502-2-2 standard.
(2) End of Life (EOL) corresponds to End of Service (EOS) as referred in the
EN45502-2-2 standard.
Shock energy
Charge time (sec)
BOL
42 J
10 2)
ERI
42 J
13 3)