User's Manual
US-ENGLISH – 55
17.2. HOLTER FUNCTION
The Holter records up to 14 tachyarrhythmia episodes as well as the
therapy history.
STORED EPISODES
PARADYM RF CRT-D 9750 stores up to 14 episodes (VF, VT, Slow VT,
SVT/ST, nonsustained).
For each episode four levels of details are presented:
─ Tachogram (to visualize PP and PR intervals)
─ Event log for the entire episode:
─ PARAD/PARAD+ analysis for each majority,
─ Delivered therapies,
─ Markers: Atrial, ventricular and biventricular markers, sensed, paced
and in relative refractory periods,
─ EGM: onset and detection of the arrhythmia, on two therapies, and the
return to slow rhythm by recording electrogram.
Therapy history: For each arrhythmia detection, each therapy
delivered (either automatically or during an electrophysiological study)
and at the end of each arrhythmia, PARADYM RF CRT-D 9750 records
the type of majority rhythm, the number of ATP sequences delivered,
the energy and the number of shocks delivered.