User's Manual

52 US-ENGLISH
Data transmitted
The data transmitted are identical to the data available during a
standard interrogation with the Orchestra Plus programmer. All
counters, histograms, IEGMs and diagnosis available in the device are
transmitted containing (not exhaustive list):
programmed parameters
Information on patient and system implanted
battery status
lead status (brady leads and defibrillation coils)
pacing counters and mean heart rate (brady)
atrial and ventricular arrhythmia counters and episodes
ventricular therapy counters
heart failure monitoring
Data are presented in the form of 2 reports to the physician: the first
one contains a summary of major counters, histograms, warnings and
diagnosis. The second one presents the 3 most important IEGM
episodes automatically selected based on the degree of severity for
the patient.
User website
On the website, the physician is able to:
consult and schedule the remote follow-ups of their patient
configure additional ways of being notified of alerts (for instance by
SMS, fax or e-mail
consult, print and export patient reports