User's Manual
US-ENGLISH – 45
15.2. MAGNET MODE
When the magnet is applied:
─ antiarrhythmia functions are inhibited (detection of rhythm
disturbances, charging, and therapy),
─ hysteresis, VV delay and AVD paced/sensed offset are set to 0,
─ pacing amplitude is set to 6 V,
─ pulse width is set to maximum,
─ pacing rate is set to the magnet rate,
─ the following functions are disabled: ventricular arrhythmia
prevention, Mode Switch, Anti-PMT, Smoothing, Rate Response.
When the magnet is removed:
─ the sensor rate is forced to the basic rate,
─ arrhythmia detection algorithms and sequential therapies are
reinitialized,
─ therapies start with the least aggressive program for each area.
The other parameters remain at their programmed value, including the
ventricular paced chamber parameter.
NOTE: The magnet is inactive during telemetry.
The magnet rate values are as follow:
Magnet rate (bpm)
96
94
91
89
87
85
Magnet period (ms)
625
641
656
672
688
703
Magnet rate (bpm)
83
82
80
78
77
Magnet period (ms)
719
734
750
766
781