User's Manual

38 US-ENGLISH
14. IMPLANT PROCEDURE
14.1. NECESSARY EQUIPMENT
Implantation of PARADYM RF CRT-D 9750 requires the following
equipment:
Sorin CRM dedicated programmer, equipped with the
SMARTVIEW software interface and with the programming head,
pacing system analyzer, as well as its sterile connecting cables, to
evaluate the pacing and sensing thresholds,
a complete set of leads with corresponding introducers,
physiological signal monitor capable of displaying simultaneously
the surface ECG and arterial pressure,
an external defibrillator with sterile external paddles,
sterile cover for the telemetry head.
14.2. PACKAGING
Contents
The PARADYM RF CRT-D 9750 and its accessories are ethylene oxide
sterilized and hermetically sealed in two-ply clear packaging meeting
international requirements.
The sterile packaging contains a defibrillator, one screwdriver, and an
insulating plug for the DF-1 defibrillation connector.
The non-sterile items contained in the outer storage package are the
implant manual, the ICD Registration Form and its envelope, the
patient booklet, the ICD ID card and 12 identification labels.
Once delivered, PARADYM RF CRT-D 9750 is programmed to as-
shipped values that are different from nominal values (see Chapter
“Programmable Parameters” for details).