User's Manual
US-ENGLISH – 37
─ Manual ATP sequences,
─ Manual shocks.
─ Rescue shock
─ Follow-up tests:
─ Pacing lead impedance,
─ Coil impedance,
─ Capacitor charge time,
─ Sensitivity test,
─ Pacing threshold tests.
─ Data storage:
─ Therapy History Report,
─ Statistics (pace/sense, therapy, shocks, and battery voltage),
─ Up to 14 complete Holter records with event logs, marker
channel notation, and electrogram records.
The connector head has five ports: atrial bipolar pace/sense, right
ventricular bipolar pace/sense, left ventricular bipolar pace and two
ports for RV and SVC defibrillation coils. Both pace/sense ports and the
pace port are compatible with the IS-1 standard and both defibrillation
ports are compatible with the DF-1 standard. Distal lead terminal
connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the
device via feedthroughs.
Programming System: The Sorin CRM programmer is used in
conjunction with specific programmer software to interrogate and
program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring: PARADYM RF CRT-D 9750 is also equipped with
the RF wireless technology which enables to remotely monitor the
patients who have the Sorin CRM SMARTVIEW Monitor installed at
home.