User's Manual

US-ENGLISH 37
Manual ATP sequences,
Manual shocks.
Rescue shock
Follow-up tests:
Pacing lead impedance,
Coil impedance,
Capacitor charge time,
Sensitivity test,
Pacing threshold tests.
Data storage:
Therapy History Report,
Statistics (pace/sense, therapy, shocks, and battery voltage),
Up to 14 complete Holter records with event logs, marker
channel notation, and electrogram records.
The connector head has five ports: atrial bipolar pace/sense, right
ventricular bipolar pace/sense, left ventricular bipolar pace and two
ports for RV and SVC defibrillation coils. Both pace/sense ports and the
pace port are compatible with the IS-1 standard and both defibrillation
ports are compatible with the DF-1 standard. Distal lead terminal
connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the
device via feedthroughs.
Programming System: The Sorin CRM programmer is used in
conjunction with specific programmer software to interrogate and
program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring: PARADYM RF CRT-D 9750 is also equipped with
the RF wireless technology which enables to remotely monitor the
patients who have the Sorin CRM SMARTVIEW Monitor installed at
home.