Therapy parameters in VT zone
ATP 1 program Values Nominal value "As shipped"
value
ATP program
OFF-Burst-Burst+Scan-Ramp-
Ramp+Scan
Burst+Scan Burst+Scan
Number of se-
quences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 3 3
Cycles in first se-
quence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 8 8
Cycles added per
sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 0 0
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95 80 80
Ramp decrement
(per cycle) (ms)
0-4-8-12-16-20-30-40-50-60 0 0
Scan decrement (per
sequence) (ms)
0-4-8-12-16-20-30-40-50-60 8 8
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4 0.5 2
Minimum cycle
length (ms)
95-110-125-140-155-170-190-205-220-
235-250-265-280-295-310
220 220
ATP 2 program Values Nominal value "As shipped"
value
ATP program
OFF-Burst-Burst+Scan-Ramp-
Ramp+Scan
Ramp Ramp
Number of se-
quences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 3 3
Cycles in first se-
quence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 6 6
Cycles added per
sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 1 1
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95 85 85
Ramp decrement
(per cycle) (ms)
0-4-8-12-16-20-30-40-50-60 8 8
Scan decrement (per
sequence) (ms)
0-4-8-12-16-20-30-40-50-60 0 0
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4 0.5 2
Minimum cycle
length (ms)
95-110-125-140-155-170-190-205-220-
235-250-265-280-295-310
220 220
21.3.2.
21. PROGRAMMABLE PARAMETERS