Table Of Contents
ATP 2 program Values Nominal value "As shipped"
value
ATP program
OFF-Burst-Burst+Scan-Ramp-
Ramp+Scan
OFF OFF
Number of se-
quences
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 3 3
Cycles in first se-
quence
1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 6 6
Cycles added per
sequence
0-1-2-3-4-5-6-7-8-9-10-11-12-13-14-15 1 1
Coupling interval (%)
50-55-60-65-70-75-80-85-90-95 85 85
Ramp decrement
(per cycle) (ms)
0-4-8-12-16-20-30-40-50-60 8 8
Scan decrement (per
sequence) (ms)
0-4-8-12-16-20-30-40-50-60 0 0
Time limit (min)
0.5-1-1.5-2-2.5-3-3.5-4 2 2
Minimum cycle
length (ms)
95-110-125-140-155-170-190-205-220-
235-250-265-280-295-310
220 220
Shock program Values Nominal value "As shipped"
value
Shock 1 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9
10-12-14-16-18-20-22-24-26-28-30-32-
34-42
OFF OFF
Shock 2 (J)
OFF-0.5-0.8-1-1.3-1.5-2-2.5-3-3.5-4-5-6-
7-8-9
10-12-14-16-18-20-22-24-26-28-30-32-
34-42
OFF OFF
Number of Max.
Shock (42 J)
OFF-1-2-3-4 OFF OFF
21. PROGRAMMABLE PARAMETERS
64 SORIN – PLATINIUM 4LV CRT-D 1744 – U904A