User's Manual

Table Of Contents
Post-shock mode Values Nominal value "As shipped"
value
Mode
OFF-VVI-DDI-DDD OFF OFF
Duration
10s-20s-30s-1min-2min-3min-4min-5min 20s 20s
Basic rate (ppm)
From 50 to 90 by steps of 5 (± 4 %) 60 60
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
19 ms)
125 125
Exercise AV delay
(ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
19 ms)
80 80
AVD Paced/Sensed
Offset (ms)
0-10-15-25-30-40-45-55-65-70-80-85-95
-100-110-115-125 1 ms)
65 65
A amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 20 %) 5 3.5
A pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
10 %)
0.35 0.35
RV amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 20 %) 5 3.5
RV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
10 %)
0.35 0.35
LV amplitude (V)
0.1-0.25-0.35 80 %)
0.5-0.6-0.75-0.85 30 %)
1-1.1-1.25-1.35-1.5-1.6-1.75-1.85-2-2.1-
2.25-2.35-2.5-2.6-2.75-2.85-3-3.1-3.25-
3.35-3.5-3.6-3.75-3.85-4-4.1-4.25-4.35-
4.5-4.6-4.75-4.85-5-5.1-5.25-5.35-5.5-
5.6-5.75-5.85-6-6.1-6.25-6.35-6.5-6.6-
6.75-6.85-7 20 %)
5 3.5
LV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1-1.1-
1.25-1.35-1.5-1.6-1.75-1.85-2 10 %)
0.35 0.35
If Multipoint LV Pacing (MP) is set to ON, LV pacing parameters (LV amplitude, LV pulse width
and LV pacing polarity) are replaced by the following parameters:
21. PROGRAMMABLE PARAMETERS
60 SORIN – PLATINIUM 4LV CRT-D 1744 – U904A