User's Manual
Table Of Contents
- 1. General description
- 2. Indications
- 3. Contraindications
- 4. Warnings and precautions
- 5. Adverse events
- 6. Clinical studies
- 7. Patient selection and treatment
- 8. Patient counselling information
- 9. Declaration of conformity
- 10. Physician guidelines
- 11. Patient information
- 12. How supplied
- 13. Device description
- 14. Implant procedure
- 14.1. Necessary equipment
- 14.2. Packaging
- 14.3. Optional equipment
- 14.4. Before opening the package
- 14.5. Prior to implantation
- 14.6. Device placement
- 14.7. Choosing the type of lead
- 14.8. Shock configuration (+ -> -)
- 14.9. Measurement of thresholds at implant
- 14.10. Leads connection
- 14.11. Device implantation
- 14.12. Tests and programming
- 15. Special modes
- 16. Main functions
- 17. Patient follow-up
- 18. Supplemental Information
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non programmable parameters
- 23. Limited warranty
- 24. Patents
- 25. Explanation of symbols
Post-shock mode Values Nominal value "As shipped"
value
Mode
OFF-VVI-DDI-DDD OFF OFF
Duration
10s-20s-30s-1min-2min-3min-4min-5min 20s 20s
Basic rate (ppm)
From 50 to 90 by steps of 5 (± 4 %) 60 60
Rest AV delay (ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
(± 19 ms)
125 125
Exercise AV delay
(ms)
30-40-45-55-65-70-80-85-95-100-110-
115-125-135-140-150-155-165-170-180-
190-195-205-210-220-225-235-250
(± 19 ms)
80 80
AVD Paced/Sensed
Offset (ms)
0-10-15-25-30-40-45-55-65-70-80-85-95
-100-110-115-125 (± 1 ms)
65 65
A amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %) 5 3.5
A pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
(± 10 %)
0.35 0.35
RV amplitude (V)
1-1.5-2-2.5-3-3.5-4-4.5-5-6 (± 20 %) 5 3.5
RV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1
(± 10 %)
0.35 0.35
LV amplitude (V)
0.1-0.25-0.35 (± 80 %)
0.5-0.6-0.75-0.85 (± 30 %)
1-1.1-1.25-1.35-1.5-1.6-1.75-1.85-2-2.1-
2.25-2.35-2.5-2.6-2.75-2.85-3-3.1-3.25-
3.35-3.5-3.6-3.75-3.85-4-4.1-4.25-4.35-
4.5-4.6-4.75-4.85-5-5.1-5.25-5.35-5.5-
5.6-5.75-5.85-6-6.1-6.25-6.35-6.5-6.6-
6.75-6.85-7 (± 20 %)
5 3.5
LV pulse width (ms)
0.12-0.25-0.35-0.5-0.6-0.75-0.85-1-1.1-
1.25-1.35-1.5-1.6-1.75-1.85-2 (± 10 %)
0.35 0.35
If Multipoint LV Pacing (MP) is set to ON, LV pacing parameters (LV amplitude, LV pulse width
and LV pacing polarity) are replaced by the following parameters:
21. PROGRAMMABLE PARAMETERS
60 SORIN – PLATINIUM 4LV CRT-D 1744 – U904A