User's Manual

Table Of Contents
Programmed
amplitude
(V)
Stored
amplitude
(V)
Mid-pulse
delivered
amplitude
(V)
2.10* 2.21 1.87
2.25* 2.34 1.98
2.35* 2.47 2.09
2.50 2.60 2.20
2.60* 2.73 2.31
2.75* 2.87 2.42
2.85* 3.00 2.53
3.00 3.15 2.67
3.10* 3.26 2.76
3.25* 3.39 2.87
3.35* 3.52 2.98
3.50 3.65 3.09
3.60* 3.78 3.20
3.75* 3.91 3.31
3.85* 4,04 3.42
4.00 4.20 3.55
4.10* 4.30 3.64
4.25* 4.43 3.75
4.35* 4.56 3.86
4.50 4.69 3.97
Programmed
amplitude
(V)
Stored
amplitude
(V)
Mid-pulse
delivered
amplitude
(V)
4.60* 4.82 4.08
4.75* 4.95 4.19
4.85* 5.08 4.30
5.00 5.25 4.44
5.10* 5.34 4.52
5.25* 5.47 4.63
5.35* 5.60 4.74
5.50* 5.73 4.85
5.60* 5.86 4.96
5.75* 5.99 5.07
5.85* 6.12 5.18
6.00 6.30 5.33
6.10* 6.38 5.40
6.25* 6.51 5.51
6.35* 6.64 5.62
6.50* 6.77 5.73
6.60* 6.90 5.84
6.75* 7.03 5.95
6.85* 7.16 6.06
7.00* 7.35 6.22
* For left ventricular amplitude only.
If Multipoint LV Pacing (MP) is set to ON, LV pacing parameters (LV amplitude, LV pulse width
and LV pacing polarity) are replaced by the following parameters:
21. PROGRAMMABLE PARAMETERS
58 SORIN – PLATINIUM 4LV CRT-D 1744 – U904A