16. MAIN FUNCTIONS
atrial and ventricular arrhythmia counters and episodes
ventricular therapy counters
Data are presented in the form of 2 reports to the physician: the first one contains a
summary of major counters, histograms, warnings and diagnosis. The second one presents
the 3 most important IEGM episodes automatically selected based on the degree of severity
for the patient.
16.7.4 . User website
On the website, the physician is able to:
consult and schedule the remote follow-ups of their patient
configure additional ways of being notified of alerts (for instance by SMS, fax or e-mail
consult, print and export patient reports
16.7.5 . Alert system
The following set of alert trigger can be independently programmed ON/OFF by the
physician using the dedicated programmer and can trigger an alert transmission:
Low or high impedance (A, RV, LV)
Abnormal coil impedance (shock lead)
Low or High shock impedance
Inefficient high energy shock
All shocks programmed OFF
Shock treated VT/VF
Lack of V pacing in CRT device
Suspicion of noise on the V lead
AT/AF occurrence
Fast V rate during AT/AF
42
SORIN INTENSIA CRT-D 174 U151A
WARNING: The use of remote monitoring does not replace regular follow-up. Therefore,
when using remote monitoring, the time period between follow-ups visits may not be
extended.