13. DEVICE DESCRIPTION
Programmer-controlled VT induction sequences,
Programmer-controlled VF inductions (30 Hz rapid pacing or shock on T),
Programmable electrogram vectors (EGM A, EGM V, RVcoil-CAN, SVC-CAN, RVcoil-
SVC, LV bip, LV tip-RV),
Real-time annotations displayed with the markers and indicating the majority rhythm,
Manual ATP sequences,
Manual shocks.
Rescue shock
Follow-up tests:
Pacing lead impedance,
Coil impedance,
Capacitor charge time,
Sensitivity test,
Pacing threshold tests.
Data storage:
Therapy History Report,
Statistics (pace/sense, therapy, shocks, and battery voltage),
Up to 14 complete Holter records with event logs, marker channel notation, and
electrogram records.
The connector head has three ports:
Atrial “IS-1” port: performs atrial bipolar pace/sense,
LV “IS-1” port: performs left ventricular bipolar pace,
RV “DF-4” port: performs right ventricular bipolar pace/sense, port for RV/SVC
defibrillation coils.
Both pace/sense ports are compatible with the IS-1 standard and defibrillation ports is
compatible with the DF-4 standard.
Distal lead terminal connections are secured with set-screws accessed via self-sealing
silicone plugs. All lead connections pass through the header into the device via
feedthroughs.
Programming System:
The Sorin programmer is used in conjunction with specific programmer software to
interrogate and program the implanted device at implant and during patient follow-up
procedures.
Remote Monitoring:
INTENSIA CRT-D 174 is also equipped with the RF wireless technology which enables to
remotely monitor the patients who have the Sorin SMARTVIEW Monitor installed at home.
SORIN INTENSIA CRT-D 174 U151A