User's Manual
13
CPR4 – UA10821A
270° and
315°
0% UT for 1
cycle
and
70% UT for
25 cycles,
50Hz
30 cycles,
60 Hz,
Monophase:
0°
270° and
315°
0% UT for 1
cycle
and
70% UT for
25 cycles,
50Hz
30 cycles,
60 Hz,
Monophase:
0°
Voltage Interruptions (IEC 61000-4-11)
0 % UT;
For 250 cy-
cles, 50 Hz
For 300 cy-
cles, 60 Hz
NA
NOTE: UT is the a.c. mains voltage prior to application of the test level.
7.6 Electromagnetic Immunity, Portable Radio Frequency Equipment
The CPR4 is intended for use in the electromagnetic environment specified below. The customer or the user of the
CPR4 should ensure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic
environment - guidance
ATTENTION: Portable RF communication devices (including peripherals such as antenna cables and external antennae)
should not be used closer to 30 cm (12 inches) to the CPR4. Otherwise, the performance of this device may be impaired.
Radiated RF IEC 61000-4-3
Proximity fields emitted by RF wireless
communication devices IEC 61000-4-3
(provisional method)
3 V/m
80 MHz to 2.7
GHz
80% MA at
1 KHz
9 V/m
710 MHz,
745 MHZ,
780 MHz,
5240 MHz,
5550 MHz,
5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz,
810 MHz,
870 MHz,
930 MHz,
1720 MHz,
1845 MHz,
1970 MHz,
2450 MHz
3 V/m
80 MHz to 2.7 GHz
80% MA at
1 KHz
9 V/m
710 MHz,
745 MHZ,
780 MHz,
5240 MHz,
5550 MHz,
5785 MHz
27 V/m
385 MHz
28 V/m
450 MHz,
810 MHz,
870 MHz,
930 MHz,
1720 MHz,
1845 MHz,
1970 MHz,
2450 MHz
Professional health care
establishment and home
care environment.
Conducted disturbances, IEC 61000-4-6
3 V
150 kHz to
80 MHz
6V in ISM
Band between
0.15 MHz -
80 MHz
80% MA to
1 KHz
3 V
150 kHz to
80 MHz
6V in ISM
Band between
0.15 MHz -
80 MHz
80% MA to
1 KHz
Professional health care
establishment and home
care environment.