User's Manual
12
CPR4 – UA10821A
Surge immunity
IEC61000-4-5
Immunity to conducted
disturbances, induced by radio-
frequency fields
IEC61000-4-6
Radiated immunity - Magnetic
fields
IEC61000-4-8
Voltage dips, short interruptions
and voltage variation immunity
IEC 61000-4-11
7.4 Electromagnetic emissions
The programmer is intended for use in the electromagnetic environment specified below. The customer or the user
of the programmer should ensure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
Electromagnetic radiation disturbance
(radiated EMISSIONS) CISPR 11
Group 1
The programmer uses RF energy only for its own
function. Its RF emissions are very low and are not
likely to cause any interference to nearby
electronic equipment.
Disturbing voltage at supply terminals con
ducted emissions CISPR 11
Class B
The programmer is suitable for use in a
professional health care establishment and home
care environment.
Harmonic current EMISSIONS IEC61000-3-2
NA
Voltage variations, voltage fluctuations and
flicker IEC61000-3-3
NA
7.5 Magnetic and Electromagnetic Immunity
The CPR4 is intended for use in the electromagnetic environment specified below. The customer or the user of the
CPR4 should ensure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic discharge (ESD)
IEC61000-4-2
± 8 kV contact
± 15 kV air
± 8 kV contact
± 15 kV air
Professional health care
establishment and home care
environment.
Electrical fast transient/burst
IEC 61000-4-4
± 2 kV for
power supply
lines
± 2 kV for
power supply
lines
Professional health care
establishment and home care
environment.
± 1 kV for
input/output
lines
± 1 kV for
input/output
lines
Surge
IEC 61000-4-5
± 1 kV
between
phases
± 2 kV
between
line(s) to earth
NA
Professional health care
establishment and home care
environment.
Magnetic field at industrial rated
frequency (IEC61000-4-8)
30 A/m
30 A/m
Professional health care
establishment and home care
environment.
Voltage dips (IEC 61000-4-11)
0% UT
For 0.5 cycle
A 0°, 45°,
90°, 135°,
180°, 225°,
0% UT
For 0.5 cy- cle
A 0°, 45°,
90°, 135°,
180°, 225°,
Professional health care
establishment and home care
environment.