User Manual
Table Of Contents
- 1. General description
- 2. Intended use and targeted population
- 3. Contraindications and adverse events
- 4. Warnings
- 5. Clinical studies
- 6. Patient selection and treatment
- 7. Patient counseling information
- 8. Declaration of conformity
- 9. Physician guidelines
- 10. Patient information
- 11. How supplied
- 12. Implant procedure
- 13. Pacemaker interrogation and upgrade
- 14. Special modes
- 15. Functions and parameters
- 16. Patient follow-up
- 17. Standby mode
- 18. Medical follow-up
- 19. Physical characteristics
- 20. Electrical characteristics
- 21. Programmable parameters
- 22. Non-programmable parameters
- 23. Warranty
- 24. Explanation of symbols
Automatic detection of implantation
Operation:
At implantation, once the ventricular lead has been connected to the pacemaker and the
case is in contact with the tissue, the device starts the detection of implantation. After five
minutes, the device automatically checks if the ventricular lead is still connected to the
pacemaker and the case still in contact with the tissue:
1. If yes, the detection of implantation is automatically confirmed.
2. If no, a new check is done by the device every five minutes, until the ventricular lead
connection and the contact between the pacemaker and the tissue are confirmed.
The automatic detection of implantation can be deactivated via the programmer. In such case
none of the automatic configuration steps described below will be done automatically; they
shall be done manually as appropriate.
Measurements of lead impedances:
Twenty minutes after the detection of implantation has been confirmed, the device starts
measuring the lead impedance(s) every 6 (six) hours and data is stored for each chamber.
Automatic launch of parameters:
Twenty minutes after the detection of implantation:
1. Memories are initiated.
2. Sensor is programmed to "Learn".
3. Statistics are reset to zero.
4. Remote monitoring is activated if programmed
5. Ventricular Autothreshold is activated if programmed
6. For the DR model, Atrial Autothreshold is activated if programmed
7. Lead polarity Switch is activated if programmed
8. For the DR model, the device switches from DDD to SafeR mode if programmed.
Autosensing
The Autosensing function allows automatic adjustment of the sensitivity for the SR model
and for both atrial and ventricular sensitivities for the DR model.
Operation:
1. If the ventricular (or atrial) Autosensing function is programmed to Auto, the ventricular
(or atrial) sensitivity is constantly adjusted to 37.5% of the mean amplitude of the
ventricular (or atrial) signals.
2. When the atrium is paced, the sensitivity tends toward 0.4 mV. For the DR model, for
the duration of an atrial arrhythmia, the atrial sensitivity tends toward 0.4 mV.
3. When the ventricle is paced, the ventricular sensitivity tends toward 1.5 mV in bipolar
configuration and toward 2.5 mV in unipolar configuration.
Programming constraints:
1. The atrial Autosensing function (Auto mode) is available only when the atrial sensing
polarity is bipolar.
15.7.1.
15.7.2.
15. FUNCTIONS AND PARAMETERS
ALIZEA – UA10414A 47