NOTE:The battery voltage can decrease before the expiration date is reached. However, the
battery voltage should be equal to or higher than 3.04V at the time of implant.
Devices MUST NOT be interrogated and programmed within the vicinity of other devices.
Also check the integrity of the sterile package. The sterility of the contents is no longer
guaranteed if the package has been pierced or altered. If the pacemaker is no longer sterile,
it should be returned in its packaging to your MicroPort representative. Any re-sterilization of
the unit is at the discretion of MicroPort.
Open the sterile package within the sterile field and carefully remove the device (see Figure -
Opening the sterile package).
1-Non-Sterile
2-Sterile
Figure - Opening the sterile package
Handling - Do not drop. If an unpackaged pacemaker is dropped onto a hard surface, return
it to MicroPort .
WARNING:
Inadequate therapy due to defective device If an unpacked device is dropped on a hard
surface during handling, electronic parts could be damaged.
Use a replacement device
Return the damaged device to MicroPort
PREPARING THE POCKET
Because of the orientation of the pacemaker connector, it is advisable to create the pocket in
a left pectoral position.
In its final position, the pacemaker should be no more than 4 - 5 cm below the skin surface.
12.5.
12. IMPLANT PROCEDURE
ALIZEA – UA10414A 31