Instruction Manual
4/8
Form 903 (03.13) ©SOR Inc.
Dimensions
Wiring 534CR-VN
Linear = mm/inches
Drawing 0090976
Dimensions are for
reference only.
Contact the factory
for certified drawings
for a particular model
number.
Three 18” flying leads are provided for connection to a terminal strip within a cabinet or a
splice within an outlet box:
Red (+)
Black (–)
Case ground (bare wire) should be connected to earth ground.
Formula for determining
maximum loop resistance:
Loop Voltage: 10 to 30 VDC; Output: 4 to 20 mA
}
R
L
(Max) =
V
Supply
- 14V
Voltmeter
1 to 5 VDC
100K ohms min.
Case Ground
1-5 Volt
Terminal
with re-
spect to
negative
Power Supply
10 to 30 VDC
Terminals shown with
transmitter at 6 o’clock