Manual
Form 1307 (04.13) ©SOR Inc. 
3/4
Calibration
This instrument may be calibrated as a pressure switch (low side vented); however, for 
best performance, calibrate under simulated operating conditions. 
The 131 has a static shift of -1 psi per 100 psi applied static. If the unit is calibrated without 
static pressure applied, the set point should be adjusted to counter the actual static pressure.
To check the set point of a switch, monitor either the common (C) and normally open (NO) 
or the common (C) and normally closed (NC) contacts for change of state. Connect the 
process connection to a regulated hydraulic or pneumatic pressure source. Monitor with 
an accurate pressure measuring standard. Slowly increase or decrease the pressure to 
accurately capture the precise moment that the switch changes state. To assure the most 
accurate and repeatable results, the switch must be tested in an identical manner each time 
the calibration is checked.
Increasing Set Points
If the normal operating pressure is below the set point, then the pressure should be increased from 0 
PSI up to the increasing set point and then back down to the reset point. Repeat this cycle 
as necessary.
Decreasing Set Points
If the normal operating pressure is above the set point, then the calibration should be checked by first 
pressurizing to the normal operating pressure, then reducing the pressure to the decreasing set point, 
and then increasing the pressure to the reset point. Repeat this cycle as necessary.
To adjust pressure at which switch will operate, remove cover and tighten the hex head adjusting nut 
with a 3/4” wrench to increase pressure; loosen to reduce pressure. 
After calibration is complete, reinstall the cover with new gaskets as required by the 
Maintenance instructions.
Switching Element
Calibration Scale
Adjusting Nut
The switching element has been 
positioned with a dial indicator to 
a tolerance of ±.002 inches. Do 
not move this switching element! 
Its position has nothing to do with 
the set point adjustment. Any 
movement can either render the 
switch inoperative or cause the 
switching element to be damaged 
with overpressure.
Do not apply more than 300 psi 
positive differential to the switch. 
Do not apply a negative differential 
of more than 10 psi.




