Warning Change the battery in a clean dry environment to prevent water/ ingress from
entering the pump case. The battery cap is correctly installed with a push and turn
counter clock-wise to locked. This prevents water/ingress.
Warning
Notice
• The Diabecare DANA-i Insulin Pump is powered by an external (AAA) Battery
Notice
• SOOIL recommend using either a Duracell alkaline AAA battery or an ‘Energiser
Advanced’ alkaline AAA battery.
Caution The pump required one AAA 1.5V battery. Use a new AAA alkaline battery. Do
not use a carbon zinc battery in your pump. Carbon zinc batteries are not compatible
with this pump.
Caution Lithium batteries are not recommended as the battery level indicator may not
be accurate.
Caution
• DO NOT attempt to change the battery while a bolus is in progress.
• Dispose of used batteries in an environmentally friendly way according to local
disposal requirements or contact your local insulin pump distributor for disposal
information.
• It is recommended to keep a spare battery as backup.
• For accurate reading of the remaining battery charge, check the battery display
following the delivery of a bolus
Caution
Caution
Caution
Insulin Pump
1. Open the battery cap by inserting the
into the battery cap slot and turning
clock-wise ¼ turn.
2. Insert the battery with the positive at
top and insert the (-) into the Insulin
Pump.
3. Replace the battery cap turning it
counter clock-wise.
4. Completed when the cap is tightly
locked with insulin pump.
2. Installing a battery
Battery Cap
With O-ring
A
A
A