User's Guide

108 109
Radiated RF
IEC 61000-4-3
10 V/m
80 to 2.7
10 V/m
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey a,
should be less than the
compliance level in each
frequency range b.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
NOTE.1 At 80 and 800 ㎒, the higher frequency range applies.
NOTE.2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption, and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Diabecare DANA-i
insulin pump is used exceeds the applicable RF compliance level above, the Diabecare
DANA-i insulin pump should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Diabecare DANA-i insulin pump.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.