User's Guide
86 87
CONFIRM PAIRING
Displayed when the pump receives a pairing signal.
NO DELIVERY
The pump cannot deliver insulin for one or more of
different reasons. This message is shown on initial
display and may blink/flash with additional information
CHECK GLUCOSE
This alarm is to remind you to check blood glucose level
after a bolus.
How to solve:
Silence the alarm by pressing any button. In this case, insulin is not delivered until prime
is properly completed. Refer to 4.7 Prime the infusion set tubing.
How to solve:
To cancel / prevent pairing, press (-) NO
Refer to instruction for use of application.
How to solve:
A detail message will blink alternately. Refer to following reason of NO DELIVERY. Refer
to instruction for use of application.
※ Reason of NO DELIVERY
Notice
The default is 2 hours and can be adjusted by a health care professional in the Dr.
Mode.
03/10/2019
245u
10:04 AM
NO DELIVERY
CONFIRM
PAIRING?
:NO- OK :YES
ALARM
CHECK
GLUCOSE
Pump has been suspended. Refer to 6.2 suspend
If there is no insulin in the reservoir, EMPTY
RESERVOIR is displayed and insulin is not injected.
Refer to 4. Loading Insulin into the Pump.
03/10/2019
0u
10:04 AM
EMPTY RESER.
03/10/2019
0u
10:04 AM
SUSPEND
If prime is not completed or was less than 7u
(minimum) to fill tube. Refer to 4.7 Prime the
infusion set tubing.
If basal setting is 0.0 u/h, NO DELIVERY is displayed
during that time.
Refer to 3.2 Setting the Basal rate.
03/10/2019
245u
10:04 AM
NO PRIME
03/10/2019
u/h 100%
245u
0.00
10:04 AM
* ALARM Message