PRIME AMOUNT IS NOT ENOUGH
This alarm/alert message will be displayed if the
volume delivered for tubing prime is less than 7 units.
How to solve:
Properly priming the infusion set tube is necessary to ensure all air is displaced and
insulin is ready for infusion. Even the shortest infusion set tube will require more than 7
units to properly prime - so for safety the Insulin Pump has a minimum required Prime
volume of 7 units. Refer to 10.3 Prime Volume of infusion sets for suggested minimum
prime amount for each infusion set.
ALARM
PRIME AMOUNT
IS NOT ENOUGH
* ALARM Message
DELIVERY LESS THAN BASAL SET RATE
If the basal is skipped and delivered less than 80% of
the basal setting, this alarm will be generated.
How to solve:
Silence the alarm by pressing any button. If you operate the pump at basal delivery
interval, the basal may occasionally skip. For a stable basal delivery, avoid long-time
pump button operation.
Basal Insulin delivery intervals very based on the size of the set basal rates.
Size of BASAL Rate (u/hr) BASAL delivery interval
0.1 U/h (Basal)
Every 4 minutes (1/15) of the hourly rate is delivered.
15 deliveries per hour.
Extended bolus
0.09 U/h (Basal)
Basal delivery will occur once at 56min the hour.
(hourly)
Warning
The individual small basal delivery increments maybe interrupted during
Bluetooth pairing or during changes being made to configuration or pump settings.
These increments of basal delivery in very low basal delivery rates such as ≤ 0.09 U/h
patients need to be monitored carefully to avoid unexpected hyperglycemia which it
could lead to ketoacidosis
ALARM
DELIVERY LESS THAN
BASAL SET RATE
How to solve:
A melody will play for 30 seconds. You can silence by depressing any button.
PRIME INCOMPLETE
If the prime process is not correctly completed
following a refill the "PRIME INCOMPLETE" alarm
occurs every 5 minutes and message will be displayed
with a beep sound.
ALARM
PRIME
INCOMPLETE