* ERROR Message
CHECK ERROR
This alarm occurs if the Insulin Pump suspects an
internal signal defect.
SYSTEM ERROR
This alarm occurs when the Insulin Pump detects any
unusual movement of the controller.
How to solve:
If/when either of these alarms, removal of the battery will silence the alert. Reinsert the
battery after 10 seconds and the pump will perform a full self-check procedure. DANA
Insulin pump is monitoring all operation for safety. Any unusual noise may cause
relevant alarms to prevent any further problems.
However, if it does not occur again after resetting the pump, the pump has no problem.
Warning
When the errors occur, all the delivery is stopped. Check the insulin delivery
following restart when these errors occur.
Caution
If WARNING persists, contact technical support from your local Insulin
Pump distributor.
WARNING
CHECK
ERROR
WARNING
SYSTEM
ERROR
* ALARM Message
SUSPEND
Select any menu related to infusion (insulin delivery)
whilst the Insulin Pump is in Suspend Mode, you are
alerted with this message.
Refer to 6.2 SUSPEND.
How to solve:
Turn the Suspend Mode off prior to continuing in any of the infusion (insulin delivery)
menu's.
ALARM
SUSPEND
* ALARM Message
MISSED BOLUS
If a bolus missed a bolus during the time period set, the
Insulin Pump will give you an alarm together with an
alarm message.
Refer to 5.3 Bolus Setting-MISSED BOLUS
How to solve:
Silence the alarm by pressing any button.
Follow the prompt by determining if a food bolus was missed and administer if
necessary.
ALARM
MISSED BOLUS
01:00PM - 02:00PM