User's Guide
66 67
1. Select BUTTON LOCK from the MAIN MENU.
2. The BUTTON LOCK symbol is displayed on
the Main Menu.
3. Press any button from the Initial Screen and a
PASSWORD request appears.
4. The Correct PASSWORD must be entered
correctly before any delivery menu can be
accessed.
3. Button Lock
Button lock prevents accidental Insulin Pump keypad presses.
It is particularly useful for:
• Pediatric patients who are not able to program their own pump.
• Patients whilst sleeping.
Notice
• The default PASSWORD is 1 2 3 4
• The PASSWORD can be changed within the USER OPTION menu.
03/10/2019
u/h 100%
245u
0.20
10:04 AM
B
MAIN MENU
BUTTON LOCK
PASSWORD
0 0 0 0
PASSWORD
1 2 3 4
MAIN MENU
BUTTON
LOCK
1. Select SUSPEND from the MAIN MENU.
2. The display changes on the initial screen to
alternate between SUSPEND and NO DELIVERY.
3. To restart delivery select SUSPEND OFF from
the main menu and confirm with . The pump
will give an alarm and the basal active status on
initial screen.
2. Suspend
To stop the Insulin Pump with the suspend function. Suspend stops all insulin delivery
including basal and bolus. The Suspend must be off to resume basal delivery or to deliver
a bolus.
Notice
When suspended, the insulin pump alarm will ring every 4 minutes. This is to advise that
no insulin is being delivered.
03/10/2019
245u
10:04 AM
SUSPEND
03/10/2019
245u
10:04 AM
NO DELIVERY
MAIN MENU
SUSPEND
MAIN MENU
SUSPEND OFF
MAIN MENU
SUSPEND