User's Guide

60 61
2. EXTENDED BOLUS
Adjust between extended and dual bolus ON/FF.
3. BOLUS CALCULATOR
Setting changes CARB, UNIT or BOTH
CARB = Bolus requests grams entered
UNIT = Bolus by adjusting units entered
BOTH = every bolus asks which option?
4. BOLUS RATE
Enables the default bolus size to be adjusted to a
personal amount.
5. MISSED BOLUS 1-4
This is a safety reminder alarm. When turned on a
time period can be set for regular meal bolus’s.
Once set an alarm will remind of a missed bolus if
no bolus is delivered during the selected time
period. Change the MISSED BOLUS to ON then
the TIME SETTING option opens.
Notice
If you do not want to use the MISSED
BOLUS feature, make sure it set to OFF”.
6. EXIT
Press , go back to BOLUS MENU.
* Bolus Setting
BOLUS SETTING
1.CIR/CF SETTING
2.EXTENDED BOLUS:OFF
3.BOLUS CAL.:CARB.
4.BOLUS RATE
5.MISSED BOLUS1:OFF
6.MISSED BOLUS2:OFF
7.MISSED BOLUS3:OFF
8.MISSED BOLUS4:OFF
9.EXIT
BOLUS SETTING
11.00 u
BOLUS
BOLUS SETTING
BOLUS SETTING
4.BOLUS RATE
5.MISSED BOLUS1:ON
6.MISSED BOLUS1:OFF
MISSED BOLUS:
08:00AM-10:30AM
BOLUS SETTING
8.MISSED BOLUS4:OFF
9.EXIT
1.CIR/CF SETTING
* Bolus Setting
1. CIR/CF SETTING
Select
CIR or CF
Press the button to adjust the CF (correction
factor) or press the button to adjust the CIR (carb
to insulin ratio).
From within the CIR or CF option, it is then
necessary to adjust the ratio for every hour 00-01,
01-02, 02-03 etc for each hour to 23-24…
After successfully changing the ratio to the
personalized requirements. Press to save the
settings.
CIR = Carbohydrate to Insulin Ratio
CIR and CF are ratio's - so they each reflect how much 1u of insulin will cover.
CIR is a setting based on the amount of carbohydrate in grams per 1u of insulin
requirement.
CF = Correction Factor
CF is a setting based on the expected change in Blood Glucose in mg/dl or mmol/L per
1u of insulin.
Notice
Follow the advice and guidance from a Healthcare Professional, Nurse or Doctor
when setting or changing CIR / CF ratio’s.
BOLUS SETTING
9.EXIT
1.CIR/CF SETTING
2.EXTENDED BOLUS:OFF
SAVE
SETTING?
:NO- OK :YES
SELECT
CIR OR CF
:CF- OK :CIR
CF SET
23-24 CF:50
00-01 CF:50
01-02 CF:50
02-03 CF:50
03-04 CF:50
04-05 CF:50
05-06 CF:50
CIR SET
23-24 CIR:20
00-01 CIR:20
01-02 CIR:20
02-03 CIR:20
03-04 CIR:20
04-05 CIR:20
05-06 CIR:20