User's Guide

42 43
6. Prime the Infusion Set Tubing
Prime every new Infusion set tube to displace air from within the tubing. Visually confirm
that all bubbles are primed from the Infusion Set tubing. Upon completion of refill process
confirm the basal is active and correct.
Warning It is important to properly prime the Tube and ensure all air is removed from
the system. The pump may not properly deliver insulin without this.
Caution PRIME is a very important process to ensure that the pump will deliver insulin
accurately. Delivery problems often result due to air within the tube and occlusion
alarms may be because of poor or insufficient PRIME. Patients are required to have
good level understanding of how to properly PRIME and why the PRIME process is
important.
Warning
Caution
Notice
Connect the infusion set tube then position the pump upright during priming for
the perfect removal of any air in the reservoir and tube.
Remained air
Remained air