User's Guide
22 23
1. Adjust the time
Setting the correct time is necessary for accurate basal insulin delivery and for retaining an
accurate record of all insulin delivery.
From within the Clock Setting menu – adjust the time using
the (+) or (-) key. Press (ok) to save the setting.
Notice
Changing 12 or 24hour clock format refer to 3.4 Setting User Options.
Notice
The Diabecare DANA-i Insulin Pump has UTC time. Setting the date and time is
only completed within the Dr Mode.
2. Setting the Basal Rate
Basal settings must be programed before using the insulin pump.
Basal insulin is required to maintain an ideal glucose level while fasting.
Basal insulin infusion rates are specific to individual patients. There are 24 hourly rates each
day, these may increase or decrease to match personal insulin resistance and other factors.
The healthcare professional will advise what the initial rates need to be set at the start.
Notice
It is only possible to EDIT the current (selected) Basal Profile. Default profile is #A.
(To change Basal Profile refer to 6.4 Basal Profile)
Step.1
Choose the Basal
Step.2
EDIT Basal Setting
OPTION
MAIN MENU
SUB MENU
CLOCK
SETTING
USER
OPTION
AIRPLANE
MODE
UTC = 0 Greenwich mean + 1 hour UK / Portugal / Europe West
-1 hour West Africa + 2 hour France / Germany / Italy
-2 hour Atlantic + 3 hour Europe East / Istanbul
-3 hour Atlantic + 4 hour Dubai
-4 hour US East + 5 hour Asia / Uzbekistan
-5 hour US Central / Chile + 6 hour India
-6 hour Canada + 7 hour Thailand
-7 hour US Pacific + 8 hour West Coast Australia / China
-8 hour Alaska + 9 hour Korea / Japan
-9 hour South Pacific Ocean +10 hour East Coast Australia
-10 hour Hawaii / Rarotonga + 11 hour Pacific / Noumea / Norfolk
-11 hour Samoa + 12 hour New Zealand
TIMEZONE : 00
UTC TIME
01/01/2019 00:00:00