User's Manual

2022. 03.
50 / 57
Appendix A. Specification
A-1. SF14L25
Complied Standard
EN/IEC 60601-1:2012
Medical electrical equipment Part 1: General
requirements for basic safety and essential performance,
BF, Non-continuous operation
EN/IEC 60601-2-37:2015
Medical Electrical Equipment Part 2-37: Particular
Requirements for the Basic Safety and Essential
Performance of Ultrasonic Medical Diagnostic and
Monitoring Equipment
EN/IEC 60601-1-2:2014
Electromagnetic Compatibility Requirement and tests,
Class A
Circumstance
Operation
Ambient Temperature
5°C ~ +35°C
Relative Humidity
25% ~ 80%, Non-Condensing
Atmospheric Pressure,
Max. Altitude
700hPa ~ 1060hPa,
3000m
Storage / Shipping
Ambient Temperature
-20°C ~ +55°C
Relative Humidity
25% ~ 93%, Non-Condensing
Atmospheric Pressure,
Max. Altitude
700hPa ~ 1060hPa,
3000m
Safety
Type of protection
Internal-Power Type, 3.8 VDC / 4,200 h
against electric shock
Type BF Applied Parts
Degree of protection
Non-continuous operation
Operating mode: 1:2 duty cycle (ON: 5 min / OFF: 10 min)
against electric shock
Hand-Held Type
Degrees of protection
IPX1