User's Manual
Rev.1.01
2022. 03.
46 / 57
Doc # : BN-SF-EOP-01
4-2
±2, ±4, ±8,
±15 kV/Air
should be at least 30%.
RF
Electromag
netic
Immunity
IEC 61000-
4-3
EN 61000-
4-3
3V/m
80 ㎒ -
2.7㎓
80% AM at
1㎑, 2㎐
The field strength outside the shielded
location from the fixed RF transmitter as
determined by electromagnetic field
survey should be less than 3V/m.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
Near Field
Immunity
for RF
Wireless
Communica
tion
Equipment
IEC 61000-
4-3
EN 61000-
4-3
Table #9 at
IEC 60601-1-
2: 2014
Electrostatic
transient
burst
IEC 61000-
4-4
EN 61000-
4-4
±2 kV, 100
k㎐
Repeated
Frequency
The quality of the mains power must be
that of a typical commercial or hospital
environment.
Surge
IEC 61000-
4-5
EN 61000-
4-5
(Wire to
Wire)
±0.5 kV, ±1
kV
The quality of the mains power must be
that of a typical commercial or hospital
environment.
Conducted
RF
IEC 61000-
4-6
EN 61000-
4-6
3V
0.15 ㎒ - 80
㎒
V in ISM
bands
between
0.15 ㎒ and
80 ㎒
80% AM at 1
The Ultrasound System is designed for
use in electromagnetic environments
where radiated RF disturbances are
controlled. Customers or users of the
ultrasonic scanner series can avoid
electromagnetic interference by
maintaining a minimum distance. It can
be used between portable and mobile RF
communication equipment (transmitter)
and ultrasonic scanner series depending
on the maximum power of the