User's Manual
Rev.1.01
2022. 03.
45 / 57
Doc # : BN-SF-EOP-01
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The Wireless Probe Series Ultrasound Scanner is intended for use in the
electromagnetic environment specified below. The customer or the user of
Wireless Probe Series Ultrasound Scanner should assure that it is used in such
environment.
Type
Emission
Test
Compliance
Electromagnetic environment -
guidance
RF
Emission
CISPR 11
EN 55011
Group1,
Class A
The Ultrasound System uses RF energy
for its internal functions only. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
RF
Emission
CISPR 11
EN 55011
Group1,
Class A
The Ultrasound System is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes.
Harmonic
Emission
IEC 61000-
3-2
EN 61000-
3-2
Class A
Flicker
Emission
IEC 61000-
3-3
EN 61000-
3-3
Pst: 1
Plt: 0.65
Tmax:0.5
dmax: 4%
dc: 3.3%
6.3.2 Electromagnetic Immunity
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The Wireless Probe Series Ultrasound Scanner is intended for use in the
electromagnetic environment specified below. The customer or the user of the Wireless
Probe Series Ultrasound Scanner should assure that it is used in such an environment.
Type
Immunity
Test
Compliance
Level
Electromagnetic environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-
4-2
EN 61000-
±8
kV/Contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity