User's Manual
Rev.1.01
2022. 03.
36 / 57
Doc # : BN-SF-EOP-01
must be checked by monitoring the echo from the needle.
① Attach the biopsy components.
② Set the system depth for the procedure to be performed and select the Biopsy
menu.
③ Immerse the probe into the water bath and insert the needle into the needle guide.
④ Confirm that the needle image is on the needle guidelines. If it is, then the needle
guide is properly aligned.
⑤ If the needle image displayed on the screen is outside of the needle guide lines,
check the needle guide.
5.4 Biopsy Specification
Probes
Biopsy
Components
Reusable/
Disposable
Needle gauge
Multi Depth
SF14L25
- Needle Guide
- Needle
Disposable
21
1, 1.5, 2 (㎝)