5. Immerse the temperature Probe in the center of the ice water bath for a period
of 40 seconds. Do not allow the probe to contact the vessel. When the self-
calibration for low temperature is complete, the screen will display the
following message:
6. Immerse the Temperature Probe in the center of the boiling water for a period
of 40 second. Do not allow the probe to contact the vessel.
When the calibration for high temperature is complete, the screen will display the
following message:
TEMPERATURE PROBE CALIBRATION
PLACE TEMPERATURE PROBE
INTO ICE WATER BATH
PLACE TEMPERATURE PROBE
INTO BOILING WATER
TEMPERATURE PROBE
CALIBRATION COMPLETED