TEMPERATURE: The temperature function prevents overheating of the sample
by continuously monitoring the sample temperature, and terminating the
ultrasonics when the temperature reaches a predetermined setpoint. The
ultrasonics is automatically reinstated when the temperature drops below the
setpoint. If the temperature of the sample must be monitored and /or controlled,
insert the optional Temperature Probe forcefully into the small jack on the rear
panel, immerse the Temperature Probe in the sample and Press the TEMP key.
The screen will display:
Using the numeric keys set the high temperature limit (setpoint).
The screen will display:
Press the ENTER/REVIEW key.
The screen will display:
REVIEW: The REVIEW function provides a “window” on the process by displaying
various operating parameters without process interruption. Pressing the
ENTER/REVIEW key repeatedly during processing will consecutively display the
following information.
a) Selected amplitude:
e.g. Amplitude Control 40%
b) Set and read temperature:
e.g. Temp Set 35°C Probe 27°C
c) Set processing time and elapsed processing time:
e.g. Set 5:30:25 Time 0:57:03
d) Selected pulsing cycle, and actual pulsing cycle:
e.g. Pulse 2.5 1.0/1.5 .5
e) Amount of power in watts, and accumulated amount of energy in JOULES delivered to the
probe:
e.g. 20 watts 0000000 Joules
f) Elapsed time since processing was initiated:
e.g. Elapsed time 1:27:33
Probe Temperature 27° C
Temperature Setpoint __° C
e.g.
Probe Temperature 27° C
Temperature Setpoint 35° C
e.g.
TIME 5:30:25 TEMP 35° C
PULSE 2.5 : 1.0 AMPL 40 %