SECTION IV – USING THE ULTRASONIC ATOMIZER
AMPLITUDE: Desired amplitude must be set in order for the Ultrasonic
Processor to be operational. AMPL. displays the amplitude selected e.g. 40%. To
set the amplitude at 40% when the ultrasonics is off, press the AMPL key and the
▲ ▼ key for a 40% reading on the screen, and then press the ENTER/REVIEW
key.
The screen will display:
The Ultrasonic Processor is now ready for continuous operation. To energize the
ultrasonics, press the START key or the footswitch. To de-energize the ultrasonics, press
the STOP key or release the footswitch.
To increase or decrease the amplitude in small increments when the ultrasonics is on,
depress the ENTER/REVIEW key twice to display AMPLITUDE CONTROL on the
screen, then depress the ▲ or ▼ key, as required.
CAUTION
Do not operate the ultrasonic power supply unless it is
connected to the converter
High voltage is present in the power supply – do not operate
with the cover removed.
AMPL 40 %
NOTE
To clear an erroneous entry press the CLEAR key.
NOTE
The START key and footswitch are mutually exclusive. If the process is initiated
by the START key, the footswitch becomes inoperative. If the process is initiated
by the footswitch, the STOP key becomes inoperative.