AMPLITUDE: The amplitude is the only parameter that must be set in order for
the Ultrasonic Processor to be operational. The other control parameters – Time
and Pulse, do not have to be set for continuous operation. AMPL. displays the
percentage of maximum of amplitude e.g. 40%, set by the AMPLITUDE control.
Rotate the AMPLITUDE control for a 40% reading on the screen – Ampl 40%.
The screen will display:
TIME __:__:__ TEMP __ __°C
PULSE__:__:__ AMPL 40 %
The Ultrasonic Processor is now ready for continuous operation. To energize ultrasonics,
press the START key or the footswitch. To de-energize ultrasonics, press the STOP key
or release the footswitch. If the Time, Temperature, Pulse, Save, or Recall functions must
be used, refer to the appropriate paragraph(s) below.
NOTE
Any combination of functions can be selected in any order. To clear an erroneous
entry press the CLEAR key.
NOTE
If the START key is pressed and the time limit has not been set, processing will remain
uninterrupted until the STOP key is depressed.
If the START key is pressed and the time limit has been set, processing will remain uninterrupted
until the set time limit expires, or the STOP key is pressed – whichever occurs first.
If a footswitch is use, and the time limit has not been set, processing will remain uninterrupted as
long as the footswitch is depressed.
If a footswitch is used, and the time limit has been set processing will remain uninterrupted until
the time limit expires or the footswitch is released – whichever occurs first.
The START key and footswitch are mutually exclusive. If the process is initiated by the START
key, the footswitch becomes inoperative. If the process is initiated by the footswitch, the STOP
key becomes inoperative.