User Manual

CAUTION
Never allow liquid to spill into the converter. Do not use the cup horn without a
splash shield
Do not allow a microtip or extender to vibrate in air for more than 10 seconds.
When working with a microtip never allow the AMPLITUDE control to be set
above the microtip limit 40%. Ignoring these instructions will cause the microtip
to fracture.
Do not allow the vibrating microtip to contact anything but the sample.
When working with low surface tension liquids, do not use a probe with a
replaceable tip.
Never energize a threaded probe without the replaceable tip, extender, or microtip
attached.
NOTE
Refer to Section IV, for general operating suggestions and ultrasonic processing
techniques.
1. Set ON/OFF power switch to ON. The switch will illuminate and the screen will
display the power rating of the Ultrasonic Processor, cautionary notices, and the
following control parameters.
TIME __:__:__ TEMP __ __°C
PULSE__:__:__ AMPL __ __%
13