Manual

1.4 Programmable Features
When you receive your Soniclean Irrigator it will be set to function automatically. The parameters are set in
the factory. The features described below may be altered as you require. The list includes the display you will
see on the machine when you are programming. To alter these parameters refer to section 6, Program Mode.
Parameter display name Descriptions
Ct Length of cleaning times
PO Ultrasonic power levels
Cn Clean time. This determines the number of cleaning cycles your machine will
perform before automatically draining. To refill, press SHIFT and hold down REFILL
for three seconds.
SP Detergent concentration
G1 Start up degassing time
G2 Degassing time prior to each clean, (pre-clean degas)
tt Test time. Set time for foil test
Id Irrigate default. Re-programs the machine to suspend the irrigator function.
Ir Irrigate rest periods. Determines the On/Off cycles of the irrigator function.
1.5 Display Messages
The following messages are displayed on the keyboard at various times during normal operation. The display
cycles through the items marked with “*” while cleaning.
Display Explanation
- - - - Together with 5 beeps, this indicates that a process has been completed. The operator
does not need to wait for the beeps to complete before starting another process.
04.58* Shows the time remaining of the clean or degas function which is currently operating.
C-01* Indicates the current cleaning cycle number.
P-03* Indicates power level selected, e.g. P-02, P-03, etc.
dGAS The machine is de-gassing.
drn The machine is draining.
drnf Bottom of the wash tank is flushed with clean water, and then drained.
FILL The machine is filling the wash tank.
OFF The machine has completed a drain or other operation and needs to be turned off at
power point.
END End of final clean.
Lid When the lid is in the open position, a flashing “Lid” message is displayed, and all
operations are paused. When the lid is lowered again the paused operations will
recommence from the point of the pause.
tESt When the lid is open to maintain ultrasonic (foil/UPM) performance test
S-2800 Irrigator Operating Manual Page 4
© Soniclean Pty Ltd Version 20 March 2014