Appendix C: Recommended S2800/S2800MRS preventative protocol
Soniclean recommended the following protocol to be carried out (daily and weekly by your authorities;
annually by your service agent) to ensure the machine is performing at the optimum performance.
DAILY
Remove and clean strainer Ensure strainer is clean.
Drain machine at end of day/session Ensure the contaminated water is not stored in tank
Check cleaning efficacy Visual examination of all load items.
IMPORTANT NOTE:
If the water stored in the machine for a long duration, water might trapped in the level sensor tube. This
might potentially cause the error in machine level sensor, and hence, the machine will not stop
automatically during water overflowing.
WEEKLY
Safety checks Check condition of lid hinges and tray
Lid interlock test Ensure safety to operator
Drainage test Ensure free-flowing drainage
Detergent dosing test Ensure the detergent can flow into the tank
ANNUALLY
Verification of calibration e.g. plumbing system, detergent intake, level sensor, electronic boxes,
MRS
Completion of all validation & performance tests
A service reminder sticker and Test Certificate will be provided after each calibration.
Where to obtain service? Return your machine to Soniclean for a full overhaul and service.
Preventative maintenance contract recommendations - please contact Soniclean for more
information.