Manual
Appendix B: S-2800 Ultrasonic Irrigator Instructions
S-2800 Irrigator Operating Manual Page 19
© 2013-2014 Soniclean Pty Ltd Version 20 March 2014
Automatic Shut Down
• When the programmed number of cleaning cycles are completed the machine will
automatically drain, and display END.
• Press and hold SHIFT & REFILL to continue cleaning or
• Turn off at power point.
Manual Shut Down
• Press SHIFT and DRAIN.
• When machine has drained display will show OFF.
• Turn off at power point.
SUMMARY OPERATING INSTRUCTIONS
• Remove tray. Turn on at power point. Machine will display 8.8.8.8, then SC4.1. Machine
will automatically drain, flush, drain the flush, fill, & degas (approx. 2+3+20=25 minutes)
• Machine will display - - - -
• Check program settings.
• The machine is now ready for use.
• Load the machine and Close the lid.
• Press CLEAN. Display shows C-01 (number of cleans), degas, 01.52 (time remaining), then
C-01 (number of cleans), P-03 (power level), 09.48 (time remaining)
• To stop cleaning at any time, just lift lid or press clean again.
• When cleaning cycle is complete, machine will display - - - -
• Lift lid and remove tray.
• When programmed number of cleans completed machine will drain and display END.