Manual

4. Operating Instructions
4.1 Prior to Operating:
Position machine on bench top or table. Ideally the lid of the machine should be at elbow height. Ensure the
machine is horizontal and the water inlet tap is turned on.
4.2 How to Operate:
4.2.1 Start Process:
1. Turn on at power point.
2. Switch on ON/OFF switch.
3. The machine will display 8.8.8.8, then SC.X.X
4. The machine will automatically drain, flush, drain the flush, fill with water and detergent and
perform a de-gassing cycle.
5. When the degassing cycle is complete, the machine is ready for use.
6. During the above cycle, the machine with display the following ( Table 1);
Display Function
8.8.8.8 Internal program
SCX.X Internal program process
Drn Draining the water.
Drnf Tank is flushing. (drain is open)
Drn Tank is draining the flush
FILL
Tank is filling with water and detergent is automatically added at the
programmed quantity. (0.5%)
De-gas The solution is degassing. This takes approximately 10 minutes.
---- The machine is ready to be loaded with instruments.
Table 1: Display during Irrigator start-up
7. It is recommended that you perform foil test daily to check the performance of your Irrigator (refer
to section 5, Foil Test), per AS2773-2, AS4187, HTM 01-05.
4.2.2 Confirm Parameters:
To check or change the settings or parameters of your Irrigator:
1. Press the “Shift” key and “Prog” key together and hold for 3 seconds.
2. A beep will sound when command has been recognised.
3. The machine will display the “Clean Time” or “ct”
4. Scroll through the parameters using the “Enter” keys
5. The parameters may be changed as required. (refer to section 6, Program Mode)
S-2800 Irrigator Operating Manual Page 9
© 2013-2014 Soniclean Pty Ltd Version 20 March 2014