Datasheet

MPI-502
Short circuit loop impedance Z in mode RCD – without RCD tripping
L-PE
Measurement using <15mA,current measurement range in accordance with IEC 61557: 0,51...1999Ω
Short circuit loop impedance measurement Z , Z , Z
L-PE L-N L-L
Measurment using 7,6/13,3A current – Measurement range according to
IEC 61557: 0,13...1999Ω (for 1,2m lead) and 0,19...1999 (for WS-01 and WS-05).Ω
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
200...1999
Ω
Ω
Ω
0,01
0,1
1
Ω
Ω
Ω
±(5% m.v. + 3 digits)
Nominal work voltage: UnL-N/ UnL-L: 220/380V, 230/400V, 240/415V,
Voltage operating range: 100…264V (for ZL-PE i ZL-N) and 100…440V (for ZL-L),
Nominal network frequency fn: 50Hz, 60Hz,
Max. test current: 23A (230V), 44A (440V),
PE terminal connection corectores check with the help of touch electrode
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
200...1999
Ω
Ω
Ω
0,01
0,1
1
Ω
Ω
Ω
±(6% m.v. + 10 digits)
±(6% m.v. + 5 digits)
test without tripping the RCD for IΔn 30mA,
nominal work voltage: Un: 220V, 230V, 240V
voltage operating range:180...270V
frequency operating range fn: 50Hz, 60Hz
PE terminal connection correctness check with the help of touch electrode
Range
Resolution
Accuracy
0,00...19,99
20,0...199,9
200...400
Ω
Ω
Ω
0,01
0,1
1
Ω
Ω
Ω
±(2% w.m. + 3 cyfry)
Low voltage test of the circuit and insulation continuity
Measurement range in accordance with
Test of PE wire continuity using a ±200mA current
IEC61557: 0,12...400Ω
„m.v” - measured value.
RCD parameters test (voltage range 180...270V):
RCD tripping test and measurment of tripping time t (for t function)
A A
Breaker Type
Test Current
Multiplier
Range Resolution
Accuracy
Standard
Selective
0,5*I
1* I
2* I
5*I
0,5*I
1* I
2* I
5*I
n
n
n
n
n
n
n
n
0...300ms
0...150ms
0...40ms
0...500ms
0...200ms
0...150ms
1ms
±(2% m.v. + 2 digits)
Range
Resolution
Accuracy
0,0V...299,9V
300V...500V
0,1V
1V
Voltage measurement
±(2% m.v. + 6 )
±(2% m.v. + 2 )
digits
digits
Frequency measurement
Range
Resolution
Accuracy
45,0...65,0Hz
0,1Hz
±(0,1% m.v. + 1 )digit
Precision of the differential current: 0,5*I :-8...0% for 1*I , 2*I , 5*I : 0...8%
Δn
Δn Δn Δn
Measurementof the RCD disconection current I for sine waveform testing current
A
Selected
Current
Range Accuracy
10mA
30mA
100mA
300mA
500mA
3,3...10,0
9,0...30,0mA
33...100mA
90...300mA
150...500mA
mA
± 5% I
n
0,1mA
1mA
It is posible to start the measurement from the positive or negative half of the forced leaking current
Selected
Current
Range
Accuracy
10mA
30mA
100mA
300mA
4,0...20,0
12,0...42,0mA
40...140mA
120...420mA
mA
±10% I
n
Resolution
Test
Current
0,1mA
1mA
0,35 x I ...2
n n
,0 x I
0,35 x I ...1
n n
,4 x I
RCD I tripping current measurement for unidirectional pulsating residual current
A
M .easurement is possible for a positive or negative forced leakage current
It is possible to start the measurement from the positive or negative half of the forced
leaking current
Test
Current
0,3 x I
n n
...1,0 x I
Resolution
MULTIFUNCTION ELECTRICAL
INSTALLATIONS METER
Voltage on open terminals: 4…9V,
Test current at R<2Ω: min. 200mA,
Autocalibration of test leads
Measurements for both polarizations of the current
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
Ω
Ω
0,01
0,1
Ω
Ω
±(5% m.v. + 5 digits) of value Z
S
Readings of fault loop impedance R and fault loop reactance X
S S
calculated and displayed for Z <200Ω.
S
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
Ω
Ω
0,01
0,1
Ω
Ω
±(6% m.v. + 10 digits) of value Z
S
Readings of fault loop impedance R and fault loop reactance X (without tripping the RCD)
S S
calculated and displayed for Z <200Ω.
S
±(6% m.v. + 5 digits) od value Z
S
Selected
Current
Range
Test
Current
Resolution
Accuracy
10mA
30mA
100mA
300mA
500mA
0,01kΩ
Earthing resistance measurement (R ) for RCD
E
4mA
12mA
40mA
120mA
200mA
0...+10% m.v. 8 digits
0...+10% m.v. ± 5 digits
±
0...+5% m.v. 5 digits±
0,01...5,00k
0,01...1,66k
1...500
1...166
1...100
Ω
Ω
Ω
Ω
Ω
1Ω
Test
Current
Resolution
Accuracy
0..9,9V
10,0..99,9V
0,1V
Touch voltage measurement referred to the rated residual current (U )
B
10% m.v. ± 5 digits
Range
0,4 x I
n
0...15% m.v.
±10% I
n
v.2
- Adapter UNI-SCHUKO (WS-05)
- Recevier - interface for radio transmission OR-1 (USB)
- operation manual
- calibration certificate
- Test lead with banana plug; 1,2m; red
- Test lead with banana plug; 1,2m; yellow
- Test lead with banana plug; 1,2m; blue
- Pin probe with banana connector; red
- Pin probe with banana connector; blue
- "Crocodile" clip K02; yellow
- Carrying case M6
- handle to suspend the meter
- hanging straps
- batteries
WAADAWS05
WAADAUSBOR1
WAPRZ1X2REBB
WAPRZ1X2YEBB
WAPRZ1X2BUBB
WASONREOGB1
WASONBUOGB1
WAKROYE20K02
WAFUTM6
WAPOZUCH1
WAPOZSZE4
-
- Test lead with banana plug 10m; red
- Test lead with banana plug 20m; red
- Probe with START button with UNI-SCHUKO (WS-01)
- Software for creation of documentation from
electrical measurementsSONEL Reports"
Test lead with banana plug 5m; red
WAPRZ005REBB
WAPRZ010REBB
WAPRZ020REBB
WAADAWS01
WAPROSONPE4
Optional accessories of meter MPI-502:
Standard accessories of meter MPI-502:
The device meets the requirements of EMC standards EN 61326-1:2006 and
EN 61326-2-2:2006.
Electric safety:
- type of insulation double, according to EN 61010-1 and IEC 61557
- measurement category IV 300V (III 600V) according to EN 61010-1
- protection class acc. to EN 60529 IP67
Other technical data:
- dimentions 220 x 98 x 58mm
- weight ~1kg
- Auto OFF time 300, 600, 900 second
- number of measurements Z and RCD (for NiMH batteries) >5000 (2 per minute)
MPI-502 is equipment to perform complete test and verify on electrical installations
according to the most common safety standards (IEC 61557, VDE 0100, BS7671).