Datasheet
MPI-502
Short circuit loop impedance Z in mode RCD – without RCD tripping
L-PE
Measurement using <15mA,current measurement range in accordance with IEC 61557: 0,51...1999Ω
Short circuit loop impedance measurement Z , Z , Z
L-PE L-N L-L
Measurment using 7,6/13,3A current – Measurement range according to
IEC 61557: 0,13...1999Ω (for 1,2m lead) and 0,19...1999 (for WS-01 and WS-05).Ω
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
200...1999
Ω
Ω
Ω
0,01
0,1
1
Ω
Ω
Ω
±(5% m.v. + 3 digits)
•
•
•
•
•
Nominal work voltage: UnL-N/ UnL-L: 220/380V, 230/400V, 240/415V,
Voltage operating range: 100…264V (for ZL-PE i ZL-N) and 100…440V (for ZL-L),
Nominal network frequency fn: 50Hz, 60Hz,
Max. test current: 23A (230V), 44A (440V),
PE terminal connection corectores check with the help of touch electrode
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
200...1999
Ω
Ω
Ω
0,01
0,1
1
Ω
Ω
Ω
±(6% m.v. + 10 digits)
±(6% m.v. + 5 digits)
•
•
•
•
•
test without tripping the RCD for IΔn ≥ 30mA,
nominal work voltage: Un: 220V, 230V, 240V
voltage operating range:180...270V
frequency operating range fn: 50Hz, 60Hz
PE terminal connection correctness check with the help of touch electrode
Range
Resolution
Accuracy
0,00...19,99
20,0...199,9
200...400
Ω
Ω
Ω
0,01
0,1
1
Ω
Ω
Ω
±(2% w.m. + 3 cyfry)
Low voltage test of the circuit and insulation continuity
Measurement range in accordance with
Test of PE wire continuity using a ±200mA current
IEC61557: 0,12...400Ω
„m.v” - measured value.
RCD parameters test (voltage range 180...270V):
RCD tripping test and measurment of tripping time t (for t function)
A A
Breaker Type
Test Current
Multiplier
Range Resolution
Accuracy
Standard
Selective
0,5*I
1* I
2* I
5*I
0,5*I
1* I
2* I
5*I
n
n
n
n
n
n
n
n
0...300ms
0...150ms
0...40ms
0...500ms
0...200ms
0...150ms
1ms
±(2% m.v. + 2 digits)
Range
Resolution
Accuracy
0,0V...299,9V
300V...500V
0,1V
1V
Voltage measurement
±(2% m.v. + 6 )
±(2% m.v. + 2 )
digits
digits
Frequency measurement
Range
Resolution
Accuracy
45,0...65,0Hz
0,1Hz
±(0,1% m.v. + 1 )digit
• Precision of the differential current: 0,5*I :-8...0% for 1*I , 2*I , 5*I : 0...8%
Δn
Δn Δn Δn
Measurementof the RCD disconection current I for sine waveform testing current
A
Selected
Current
Range Accuracy
10mA
30mA
100mA
300mA
500mA
3,3...10,0
9,0...30,0mA
33...100mA
90...300mA
150...500mA
mA
± 5% I
n
0,1mA
1mA
• It is posible to start the measurement from the positive or negative half of the forced leaking current
Selected
Current
Range
Accuracy
10mA
30mA
100mA
300mA
4,0...20,0
12,0...42,0mA
40...140mA
120...420mA
mA
±10% I
n
Resolution
Test
Current
0,1mA
1mA
0,35 x I ...2
n n
,0 x I
0,35 x I ...1
n n
,4 x I
RCD I tripping current measurement for unidirectional pulsating residual current
A
• M .easurement is possible for a positive or negative forced leakage current
It is possible to start the measurement from the positive or negative half of the forced
leaking current
Test
Current
0,3 x I
n n
...1,0 x I
Resolution
MULTIFUNCTION ELECTRICAL
INSTALLATIONS METER
•
•
•
•
Voltage on open terminals: 4…9V,
Test current at R<2Ω: min. 200mA,
Autocalibration of test leads
Measurements for both polarizations of the current
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
Ω
Ω
0,01
0,1
Ω
Ω
±(5% m.v. + 5 digits) of value Z
S
Readings of fault loop impedance R and fault loop reactance X
S S
calculated and displayed for Z <200Ω.
S
•
Display range
Resolution
Accuracy
0,00...19,99
20,0...199,9
Ω
Ω
0,01
0,1
Ω
Ω
±(6% m.v. + 10 digits) of value Z
S
Readings of fault loop impedance R and fault loop reactance X (without tripping the RCD)
S S
calculated and displayed for Z <200Ω.
S
•
±(6% m.v. + 5 digits) od value Z
S
Selected
Current
Range
Test
Current
Resolution
Accuracy
10mA
30mA
100mA
300mA
500mA
0,01kΩ
Earthing resistance measurement (R ) for RCD
E
4mA
12mA
40mA
120mA
200mA
0...+10% m.v. 8 digits
0...+10% m.v. ± 5 digits
±
0...+5% m.v. 5 digits±
0,01...5,00k
0,01...1,66k
1...500
1...166
1...100
Ω
Ω
Ω
Ω
Ω
1Ω
Test
Current
Resolution
Accuracy
0..9,9V
10,0..99,9V
0,1V
Touch voltage measurement referred to the rated residual current (U )
B
10% m.v. ± 5 digits
Range
0,4 x I
n
0...15% m.v.
±10% I
n
v.2
- Adapter UNI-SCHUKO (WS-05)
- Recevier - interface for radio transmission OR-1 (USB)
- operation manual
- calibration certificate
- Test lead with banana plug; 1,2m; red
- Test lead with banana plug; 1,2m; yellow
- Test lead with banana plug; 1,2m; blue
- Pin probe with banana connector; red
- Pin probe with banana connector; blue
- "Crocodile" clip K02; yellow
- Carrying case M6
- handle to suspend the meter
- hanging straps
- batteries
WAADAWS05
WAADAUSBOR1
WAPRZ1X2REBB
WAPRZ1X2YEBB
WAPRZ1X2BUBB
WASONREOGB1
WASONBUOGB1
WAKROYE20K02
WAFUTM6
WAPOZUCH1
WAPOZSZE4
-
- Test lead with banana plug 10m; red
- Test lead with banana plug 20m; red
- Probe with START button with UNI-SCHUKO (WS-01)
- Software for creation of documentation from
electrical measurements “SONEL Reports"
Test lead with banana plug 5m; red
WAPRZ005REBB
WAPRZ010REBB
WAPRZ020REBB
WAADAWS01
WAPROSONPE4
Optional accessories of meter MPI-502:
Standard accessories of meter MPI-502:
The device meets the requirements of EMC standards EN 61326-1:2006 and
EN 61326-2-2:2006.
Electric safety:
- type of insulation double, according to EN 61010-1 and IEC 61557
- measurement category IV 300V (III 600V) according to EN 61010-1
- protection class acc. to EN 60529 IP67
Other technical data:
- dimentions 220 x 98 x 58mm
- weight ~1kg
- Auto OFF time 300, 600, 900 second
- number of measurements Z and RCD (for NiMH batteries) >5000 (2 per minute)
MPI-502 is equipment to perform complete test and verify on electrical installations
according to the most common safety standards (IEC 61557, VDE 0100, BS7671).