Datasheet
v.3
Electric security
- EN 61010-1 and IEC 61557
- measurement category CAT IV 600V acc. to EN 61010-1
- protection class acc. to EN 60529 IP65
Other technical data:
- power supply batteries LR03 (AAA) (2 pcs.) or rechargeable battery
- measurement current R 1,2mA ± 0,2mA
ISO
- max interference voltage at which a measurement is realised R 20V
ISO
- number of R measurements > 500
ISO
- dimensions approx. 240 x 60 x 30 mm
- weight (including batteries) approx. 0,3kg
- auto power OFF 5 min.
- electromagnetic compatibility according to EN 61000-6-3 and EN 61000-6-2
- quality standard ISO 9001
:
type of insulation double, according to
Low voltage resistance measurement in range
uitable for
acklit display
meets the requirements of the standard
Insulation resistance measurement
- the measurement voltage is selected between the two values 250V or 500V,
- on-line insulation resistance measurement,
- automatic discharge of tested object’s capacitance after measurement
of insulation resistance,
- acoustic determination of 5 seconds time intervals that facilitates to take time
characteristics during insulation resistance measurement,
- protection against live objects.
0...2000Ω:
- current measurement <10mA with resolution 0.1Ω,
- quick acoustic signalling (beeper) for circuit resistance below 10Ω.
AC/DC voltage measurement in range 0...600V:
- automatic detection AC/DC voltage.
Autoranging.
S 50Hz and 60Hz network.
B .
„AUTO-OFF” function.
Meter EN 61557.
:
U =250V
ISO
Measurement range: IEC 61557-2: 250kΩ...1000MΩ
Range Resolution
Accuracy
1...249kΩ
250...1999kΩ
2,00...19.99MΩ
20,0...199.9MΩ
200...1000MΩ
1k
1kΩ
0.01MΩ
0.1MΩ
1MΩ
Ω
not specified
±(3% m.v. + 8 digits)
U =500V
ISO
Measurement range: IEC 61557-2: 500kΩ...1999MΩ
Range Resolution
Accuracy
1...499k
500...1999kΩ
2,00...19.99MΩ
20,0...199.9MΩ
200...1999MΩ
Ω
1k
1kΩ
0.01MΩ
0.1MΩ
1MΩ
Ω
not specified
Resistance measurement:
Range Resolution
Accuracy
0,0...199.9
200...1999Ω
Ω 0.1
1Ω
Ω
±(4% + digits)m.v. 3
ontinuous audible signal for
easurement current (with open terminals for U > 3.0V): < 10mA
bat
ax voltage on open terminals
ax voltage for which measurement is still possible
c R < 10
m
m : 4...24V
m interference : +7V/-1V DC, 5V AC
Ω
±(3% + 8 digits)m.v.
Rated operational conditions:
- operating temperature 0...+40°C
- reference temperature 23 ± 2°C
- storage temperature -20...+70°C
MIC-2
Insulation resistance measurement:
AC/DC voltage measurement
Range Resolution
Accuracy
0...600V 1V
±(3% + 2 digits)m.v.
„m.v.” - measured value.
Directive: Low Voltage Directive (LVD) 2005/95/WE.
Standards: EN 61010-1:2001 Safety requirements for electrical equipment for
measurement, control and laboratory use. General requirements.
EN 61010-031:2002/A1:2008: Safety requirements for electrical
equipment for measurement, control and laboratory use. Safety
requirements for hand-held probe assemblies for electrical
measurement and test.
EN 61557-2:2007 Electrical safety in low-voltage distribution systems
up to 1000 V AC and 1500 V DC - Equipment for testing, measuring,
or monitoring of protective measures Part 2: Insulation resistance.
This part of IEC 61557 specifies the requirements applicable to
equipment for measuring the insulation resistance of equipment and
installations in the de-energized state.
Directive: Electromagnetic compatibility (EMC): 2004/108/WE.
Standards: EN 61325-1:2006 Electrical equipment for measurement, control and
laboratory use - EMC requirements.
The group of Insulation Resistance Meters conforms the following directives or
standards according to CE requirements.