Instructions
Well Casing & Depth Indicator Probe Replacement
Model 105
®
Solinst is a registered trademark of Solinst Canada Ltd.
Solinst Canada Ltd. 35 Todd Road, Georgetown, Ontario Canada L7G 4R8 www.solinst.com
E-mail: instruments@solinst.com Tel: +1 (905) 873-2255; (800) 661-2023 Fax: +1 (905) 873-1992
May 2, 2022
(# 116339)
Instructions
1. Make sure the Well Casing & Depth Indicator is turned off.
2. To remove the old probe, twist the probe counter-clockwise
and slowly pull it away from the tape seal plug.
3. It will take some force to pull past the o-rings on the tape
seal plug. Ensure not to overextend the wire connectors
inside the probe.
8. If required, use tweezers to carefully pull the green and
white wire connectors from the replacement probe body.
9. To attach the new probe, connect the green wire to the
brass connector on the top of the tape seal plug and the
white wire to the bottom brass tube. Ensure the connectors
are pushed all the way onto the brass tubes.
10. Keeping the wires connected, slowly twist the probe up to
two times. This shortens the wires and allows them to tuck
easier into the probe body.
Tools and Materials Needed
1. 105 Replacement Probe Assembly (#114742)
• Includes two replacement #V013 O-Rings
2. Tweezers (if required)
Note: To avoid any nicks, make sure the wires are tucked back into the
probe body when pushing the probe onto the tape seal plug.
11. Line up the indents in the probe with the grooves in the
tape seal plug. Push the probe past the o-rings, then twist
the probe clockwise until the probe seats on the tape seal
plug.
12. Turn the Well Casing & Depth Indicator on. Press the
plunger at the bottom of the probe into the probe body.
If the connections are correct, the buzzer and light will be
activated. Check the connections if the Indicator does not
operate as expected.
Probe Replacement Video
http://bit.ly/2M812LM
Note: To avoid damage to the tape seal plug, do not use a sharp tool to
remove or install the o-rings.
4. Remove the two connectors from the brass tubes.
5. Remove the old o-rings. Use your fingers to push the
o-rings up and roll it off the tape seal plug.
6. Use your fingers to install the new o-rings in the grooves.
You can lubricate them with a bit of water if required.
7. Lay the tape and tape seal plug so the numbers on the tape
are facing up.
Green Wire#V013 O-ring
Brass
Tubes
White Wire
Replacement Probe
Tape Seal Plug
Top of the
Numbers
Bottom of the
Numbers
Green Wire
Brass
Tubes
White Wire
Replacement Probe
Tape Seal Plug
#V013 O-rings
More Info | Instructions | Get Quote