7. Calibration and adjustment
Calibration stage
based on position 0°
Place the DL/DLM with the measurement surface horizontally
Press the
and
buttons briey
Display -1- appears on the screen
Press the
button briey, display -1- ashes for
approx. 5 seconds
Display -2- appears on the screen
Turn DL/DLM by 180°
Press the
button briey, display -2- ashes for approx.
5 seconds
Measurement display appears on the screen, calibration stage
1 is complete.
Calibration stage
based on position 180°
Turn DL/DLM „upside down“ and place horizontally
Press the
and buttons briey
Display -1- appears on the screen
Press the button briey, display -1- ashes for approx.
5 seconds
Display -2- appears on the screen
Turn DL/DLM again by 180°
Press the button briey, display -2- ashes for approx.
5 seconds
Measurement display appears on the screen, calibration stage
2 is complete.
Calibration stage
based on position 90°
Place the DL/DLM with the measurement surface vertically
Press the and buttons briey
Display -1- appears on the screen
Press the button briey, display -1- ashes for approx. 5 seconds
Display -2- appears on the screen
Turn DL/DLM by 180°