Electromagnetic Emissions and Immunity Declarations
Guidance and Manufacturer’s Declaration—Electromagnetic Emissions
The CADD®-Solis pump is intended for use in the electromagnetic environment specified below. The customer or the user of
the CADD®-Solis pump should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment—Guidance
RF emissions
CISPR 11
Group 1 The Pump uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class B The Pump is suitable for use in all establishments, including
domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Not applicable
Compliance using: 100240 VAC 50/60 Hz to 7 VDC switching AC adapter, with an AC power cord length of
1.8 m (6 ft); rechargeable battery pack; remote dose cord with a length of 152 cm ± 5 cm (60 in ± 2 in); and a
USB cable length of less than 2 m (6.5 ft).
WARNING:
The use of power supplies and a remote dose cord other than those listed in the electromagnetic emissions
declaration may result in increased emissions or decreased immunity of the pump.
The pump should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, you should verify normal operation of the pump in the configuration in which it is to be used.
Common portable and mobile consumer electronic devices may cause interference with the pump.
Observe the pump to verify normal operation.
Facility wiring must comply with all applicable electrical codes. Do not bypass power cord connections. Do
not remove a prong from the power cord. Failure to comply may result in fire or electrical shock.
Guidance and Manufacturer’s Declaration —Electromagnetic Immunity
The CADD®-Solis pump is intended for use in the electromagnetic environment specified below. The customer or user of the
CADD®-Solis pump should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance Level Electromagnetic Environment
Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 15 kV air
(IEC 60601-2-24)
Floors should be wood, concrete, or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV
for power supply lines
± 1 kV
for input/output lines
± 2 kV for power supply
lines
Not applicable
Mains power quality should be that of a
typical commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input lines
IEC 61000-4-11
< 5% Ut (> 95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
< 5% Ut (> 95% dip in Ut)
for 5 sec
< 5% Ut (> 95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
< 5% Ut (> 95 % dip in
Ut) for 5 sec
Mains power quality should be that of a
typical commercial or hospital environment.
If the user of the Pump requires
continued operation during power mains
interruptions, it is recommended that the
Pump be powered from an uninterruptible
power supply or a battery.
Power frequency
50/60 Hz magnetic field
IEC 61000-4-8
3 A/m 400 A/m
(IEC 60601-2-24)
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Note: Ut is the a.c. mains voltage prior to application of the test level.