Operator's Manual
77
RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Administrator Settings Specifications
KVO rate • 0 mL/hr if continuous rate is 0 mL/hr
• 0.1 mL/hr if continuous rate > 0 mL/hr, with standard administration set
• 0.2 mL/hr if continuous rate > 0 mL/hr, with high volume administration set
Bolus interval type • Bolus interval
• PCA lockout
Maximum delivery
rate (PCA dose,
clinician bolus,
intermittent bolus)
• With standard administration set: 250 mL/hr
• With high volume administration set: 500 mL/hr
• Max. delivery rate = continuous rate + bolus rate (PCA dose or clinician bolus
or intermittent bolus)
• Boluses may not be delivered simultaneously
Delivery limit method • Delivery limit
• Max doses per hour
• Not in use
Delivery limit period 1 to 12 hours in increments of 1 hour
Pump stopped
alarm
• Informational
• High priority
Res vol low trip
point
1 to 999 mL in increments of 1 mL
Res vol empty alarm • Insistent and one time only
• Non-insistent and repeating
Air detector • On
• Off
Air detector
sensitivity
• High
• Low
Upstream occlusion
sensor
• On
• Off
Note: The upstream occlusion sensor is automatically disabled during use
with medication cassette reservoirs.
Downstream
occlusion sensitivity
High Sensitivity: When the high pressure alarm threshold is reached, the
downstream occlusion alarm is triggered immediately.
Low Sensitivity: When the high pressure alarm threshold is reached, the
downstream occlusion alarm is delayed for 2 seconds. This allows for the
pressure to stabilize before a possible alarm. If the pressure stabilizes below
the high pressure alarm threshold before the 2 second delay is complete,
the alarm will not occur.
PM reminder Interval: 1 to 24 months in 1 month increments
Enable: On or Off
Custom keypad code 001 to 899 in increments of 1
Custom clinician
code
001 to 899 in increments of 1
Custom admin. code 001 to 899 in increments of 1
Time format • 00:00 to 23:59 military
• 12-hour am/pm
Date format • US standard (month/day/year)
• European standard (day/month/year)
• International standard ISO 8601:2004 (year/month/day)
Alarm volume • High
• Medium
• Low
Key beep • On
• Off