Operator's Manual

77
RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Administrator Settings Specifications
KVO rate 0 mL/hr if continuous rate is 0 mL/hr
0.1 mL/hr if continuous rate > 0 mL/hr, with standard administration set
0.2 mL/hr if continuous rate > 0 mL/hr, with high volume administration set
Bolus interval type Bolus interval
PCA lockout
Maximum delivery
rate (PCA dose,
clinician bolus,
intermittent bolus)
With standard administration set: 250 mL/hr
With high volume administration set: 500 mL/hr
Max. delivery rate = continuous rate + bolus rate (PCA dose or clinician bolus
or intermittent bolus)
Boluses may not be delivered simultaneously
Delivery limit method Delivery limit
Max doses per hour
Not in use
Delivery limit period 1 to 12 hours in increments of 1 hour
Pump stopped
alarm
Informational
High priority
Res vol low trip
point
1 to 999 mL in increments of 1 mL
Res vol empty alarm Insistent and one time only
Non-insistent and repeating
Air detector On
Off
Air detector
sensitivity
High
Low
Upstream occlusion
sensor
On
Off
Note: The upstream occlusion sensor is automatically disabled during use
with medication cassette reservoirs.
Downstream
occlusion sensitivity
High Sensitivity: When the high pressure alarm threshold is reached, the
downstream occlusion alarm is triggered immediately.
Low Sensitivity: When the high pressure alarm threshold is reached, the
downstream occlusion alarm is delayed for 2 seconds. This allows for the
pressure to stabilize before a possible alarm. If the pressure stabilizes below
the high pressure alarm threshold before the 2 second delay is complete,
the alarm will not occur.
PM reminder Interval: 1 to 24 months in 1 month increments
Enable: On or Off
Custom keypad code 001 to 899 in increments of 1
Custom clinician
code
001 to 899 in increments of 1
Custom admin. code 001 to 899 in increments of 1
Time format 00:00 to 23:59 military
12-hour am/pm
Date format US standard (month/day/year)
European standard (day/month/year)
International standard ISO 8601:2004 (year/month/day)
Alarm volume High
Medium
Low
Key beep On
Off