RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Maximum time to
occlusion alarm
and
Maximum Bolus
volume at
occlusion alarm
The pressure build-up that occurs after an occlusion may cause fluid to
accumulate or be stored in the line. This extra fluid may be delivered as a
bolus when the occlusion is released. For corrective action, see the steps listed
under “Downstream occlusion” in the Alarms and Messages, Alphabetical List on
page 58.
Flow
Rate
(mL/hr)
Tubing Set Max. Time to
Occlusion
Max. Bolus at
Occlusion
Spec.
(min)
Spec.
(mL)
0.1
CADD™
medication cassette
reservoir < 21-7002 with CADD®
extension set < 21-7047
≤ 160 ≤ 0.25
CADD® administration set
< 21-7091
≤ 190 ≤ 0.30
CADD® high volume
administration set < 21-7055
≤ 1200 ≤ 1.40
Flow
Rate
(mL/hr)
Tubing Set Max. Time to
Occlusion
Max. Bolus at
Occlusion
Spec.
(sec)
Spec.
(mL)
10
CADD™
medication cassette
reservoir < 21-7002 with CADD®
extension set < 21-7047
≤ 60 ≤ 0.25
CADD® administration set
< 21-7091
≤ 150 ≤ 0.30
CADD® high volume
administration set < 21-7055
≤ 580 ≤ 1.40