Operator's Manual

72
RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Rechargeable
battery pack
life with screen
backlight intensity
set to 3
These estimates are based on laboratory tests conducted at room temperature
using a new CADD®-Solis rechargeable battery pack. Actual battery life varies
depending on the temperature conditions, delivery rate, and frequency of
screen display and backlighting. It is recommended that new batteries be kept
available for replacement.
Continuous Delivery
Delivery Rate
(mL/hr)
Operating Time
(hr)
Volume Delivered
(mL)
0.4 74 29
1. 67 67
5.0 60 300
10.0 50 500
30.0 40 1200
Intermittent Bolus Delivery
IB Volume
(mL)
IB Interval
(min)
Maximum
Delivery Rate
(mL/hr)
Operating
Time
(hrs)
Volume
Delivered (mL)
5 30 40 71 710
10 30 250 60 1200
10 30 500 58 1160
Pump alarms High priority alarms: Air in line detected, Battery depleted while delivering,
Battery removed while delivering, Battery unusable while delivering,
Disposable attached improperly, Disposable damaged, Disposable detached
while delivering, Disposable locked but not latched, Disposable type high flow
administration set not allowed, Disposable type high flow administration set
required, Disposable type invalid, Downstream occlusion, Key stuck, Pressure
sensor faulty, Pump automatically stopped, Rechargeable battery end of life,
Remote dose cord key stuck, Reservoir volume empty, Stop mode reminder,
Upstream occlusion, wireless communication module intermittent connection
Medium priority alarms: 19
Low priority alarms: 8
Informational messages/alerts 23
Delivery rate
during priming
Standard volume cassette tubing: approx. 250 mL/hr
High flow volume cassette tubing: approx 500 mL/hr
Alarm disabled
during priming
Air-In Line
Maximum
infusion pressure
1.86 bar
27.0 p si
High pressure
alarm threshold
1.24 bar ± 0.62 bar
18 ± 9 psi
Air detector
alarm
High Sensitivity: Single bubble > 150 μL
Accumulated air > 2mL over 15 minutes
Low Sensitivity: Single bubble > 2mL (2000 μL)
Accumulated air > 4mL over 15 minutes
Note: Bubbles smaller than 50 μL are not included in the accumulated air measurement.
Statistical tolerance limits calculated at 95/99 level.
Maximum
volume infused
under single-
fault conditions
CADD
®
administration set: 0.15 mL
CADD
®
high volume administration set: 0.30 mL