RefeRences and TRoubleshooTing
RefeRences and
TRoubleshooTing
Battery status Battery State CAD-Solis Pump Status
25% to 100% No alarm
Low battery Transition to low battery condition
Battery low message appears
Pump emits 3 beeps every 5 min
Low battery warning message appears on pump display
Pump is operable
LCD backlight flashes for 12 ms during each motor operation
Depleted
battery
Transition to depleted battery condition
Battery depleted message appears
Pump emits a continuous, variable-tone alarm
Depleted battery warning message appears on pump display
Battery power is too low to operate pump
Pump delivery operation stops
Shut down Pump shuts off due to low operating voltage
Alkaline battery
life with screen
backlight intensity
set to 3
These estimates are based on laboratory tests conducted at room temperature
using new batteries (Duracell
®
PC1500 / MN1500, IEC LR6). Actual battery life
varies depending on the battery brand, shelf life, temperature conditions,
delivery rate, and frequency of screen display and backlighting. It is
recommended that new batteries be kept available for replacement.
Continuous Delivery
Delivery Rate
(mL/hr)
Operating Time
(hr)
Volume Delivered
(mL)
0.4 142 56
1. 139 139
5.0 124 620
10.0 113 1130
30.0 69 2070
Intermittent Bolus Delivery
IB Volume
(mL)
IB Interval
(min)
Maximum
Delivery Rate
(mL/hr)
Operating
Time
(hrs)
Volume
Delivered
(mL)
5 30 40 90 900
10 30 250 104 2080
10 30 500 88 1760